Range and Model Information
If your Dosimeter is going to be used in harsh conditions where scratching
of the lens is a concern we recommend a
synthetic
sapphire window to protect the lens
even in the toughest of conditions.
Arrow-Tech,
Inc. recommends a yearly calibration of the Direct
Reading
Dosimeter which is generally consistent with good health physics practices. More
frequent calibration may be necessary should the user’s license require a
shorter calibration interval.
|
How does the
Dosimeter work? |
|
Reading a Dosimeter
The way we "read" a dosimeter is to just
point it at a light source so that we can look through it and see the
scale. The conductive fiber moves across a very clear well marked
scale that produces the reading. |
|
 |
| Arrow-Tech dosimeters use an extremely sensitive fiber electrometer type
voltmeter and a small volume of air to measure the total amount of radiation to which the
instrument has been exposed. A reading may be made at any time by merely looking at a
source of light through the eyepiece end of the instrument. The reading
here is at zero. This particular dosimeter is the 200 mR model.
|
|
 |
|
|
The graduated scale shown is what is viewed
when looking through the Dosimeter
|
| Wearing a
Dosimeter
You can wear the dosimeter like a pen in your pocket, or
clip it to your belt. You can also set it outside for an hour, this
will tell you the radiation rate. You don't have to have a radiation
meter if you have a dosimeter and are not in a hurry to measure the rate. |
|
|
| Also, if ten or a hundred people are together or
going somewhere, only one or two of them need to wear a dosimeter.
So long as they all stay together, the radiation dosage that one gets will
be what they all get. It is nice to have two or three, so that you
can use one to check the other and make sure that they are working. |
| The dosimeter is charged (reset to zero) so the
dosimeter can be used over and over again. This is usually
done before each use.
Click here for instructions on charging
the dosimeter. |

BUY NOW..
Specifications
| Radiation Detected: |
Gamma and X-ray from 16 KeV to 6 MeV |
| Ranges: |
0 - 200 mR to 0 - 600 R |
| Energy Response: |
See Response curve below |
| Detector: |
Fiber electrometer mounted in an electrically conducting plastic
ion chamber |
| Detector Housing: |
Very low permeability plastics - hermetically sealed |
| Accuracy: |
Within + or - 10% of true exposure |
| Rate Response: |
Dose rate independent for gamma and x-radiation |
| Electrical Leakage: |
Less than 0.5% of full scale for 24 hours at 50 degrees C |
| Temperature Range: |
-20 degrees C to 50 degrees C |
| Relative Humidity: |
Up to 90% |
| Dimensions: |
Length 4.5 in. (12.4cm), Diameter: 0.6in (1.5cm) |
| Weight: |
1.0 oz. (25 grams) |
| Finish: |
Barrell and end caps: Natural matte black
Clips: Metal clips |
| Warranty: |
2 year limited warranty |
| ANSI Spec: |
Manufactured to ANSI N322-1997 & N13.5 |
Manufactured in the United States
Superior Energy Response: 16 KeV to 6 MeV
Rugged: Meets ANSI Specifications N13.5 and N322-1997
Low Electrical Leakage
Low Range models can measure Background Radiation