Radiation Detection
Area Monitors Portable
- AMP 50
PDF
file AMP - 100
AMP - 200
|
The
AMP-50 is a GM tube-based low
range Area Monitor. It has
been designed specifically to be used in low dose rate fields using a
very sensitive GM tube.
The
AMP-50’s detector features a linear response from 10 µR/h to 4 R/h
(0.1 µSv/h to 40 mSv/h ).
|
The
probe head may be located near a filter cube, rad waste stream, resin
tank, or even inside the fuel pool (to take advantage of waterproof
characteristics). The AMP-50 may be used in one of 2 ways: by locally
reading the smoothed digital display via the hand-held meter, or by
connecting the meter to a Remote
Monitoring System (e.g.
wired DDC 16 or wireless WRMPlus) and TeleMap or connected directly to
RS-232 computer input (using the RS-232 cable)
|

|
|
Features |
Wide
range response from 10 µR/h to 4 R/h (0.1 µSv/h to 40 mSv/h).
Ruggedized construction, waterproof detector housing and cable
Quick-connect connectors allow customization of cable length and
facilitate easy de-contamination
Built-in RS-232 communication connection for use with Area Monitor or
WRM transmitter
"Smoothed" digital display offers accurate, stable readings
User-selectable internal alarm threshold |
|
Applications |
Area
monitoring of low dose rate areas (ISFSI -Independent Spent Fuel Storage
Installations)
Low-level radioactive waste material surveillance
Facility & site boundary surveillance
Emergency Response surveillance
Control shipping of radioactive material
Environmental monitoring
Outdoor area monitoring
Underwater surveys
Alara waiting areas |
|
Technical
Description |
|
The
Area Monitor Probe (AMP-50) is a low-range GM tube-based detector
designed to be continuously used in areas where exposure levels are
required to be monitored.
The
instrument consists of three parts: the Meter box, which includes the
display unit, operating pushbuttons and detector’s electronics; The
probe head, which contains the GM Tube; and the connecting cable,
which is fitted with quick-connect-type connectors at each end.
The AMP-50’s connections and probe head feature watertight
sealing to allow for use in underwater applications.
|
|
Specifications |
|
Detector:
|
|
(ZP-1201,
or equivalent)
|
|
Energy
Range:
|
|
50
KeV to 2 MeV.
|
|
Measuring
Range:
|
|
10µ
R/h to 4 R/h (0.1 µSv/h
to 40 mSv/h)
|
|
Sensitivity
(137Cs):
|
|
17
cps per mR/hr
|
|
Accuracy:
|
|
±10%
of reading within measuring range
|
|
Units:
|
|
mR/h
and mR or m Sv/h and m Sv
|
|
Controls:
|
|
Four
key keypad with positive feedback. Key combinations for
programming operating parameters.
|
|
Power
Source:
|
|
One
9-volt cell battery or external 9V power supply, 50 hours
minimum continuous operation, using an alkaline battery
(speaker off)
Automatic battery check under full load (without backlighting)
|
|
Display
:
|
|
LCD
Display showing: Detector failure, Low battery, Overflow,
Threshold
|
|
Temperature
Range:
|
|
Operation:
-10° C to +50° C (15° F to 122° F)
Storage: -20° C to +60° C (-5° F to 140° F)
|
|
Humidity
Range
|
|
40%
to 95% RH (non condensing)
|
|
Casing
Material:
|
|
Meter:
Aluminum
Detector: Aluminum, waterproof
|
|
Dimensions:
|
|
Meter:
2.7” wide, 4.7” high, 1.25”
deep
Detector: Length:6.2”,
diameter:1.3”
|
|
Weight:
|
|
Meter:
340 g (0.76 lbs) including battery
Detector: 131 g (0.29 lbs) without cable
|
|
Std.
Cable Length:
|
|
25
ft.
|
|
Max.
Cable Length:
|
|
350
ft.
|
|
Expected
detector lifetime:
|
|
250
h at
1000 R/h
|
Accuracy,
range and sensitivity details are related to 137Cs |
|
Ordering
Number:_
#4-0026 |
|
|